A Comprehe nsive Guide to Xenon Arc Tester vs. UV Weathering Tester
This guide will walk you through everything you need to know—just like a practical industry handbook.
This Color Fastness to Light Tester apply to light color fastness, anti-climate color fastness, light & perspiration color fastness and light aging tests of textile, paint, dye, coating, rubber, plastic, wood floor, paper and other materials, it have anti-climate functions of exposing to the rain, revolution and mist spraying.
This instrument is also widely used as an ISO 105 B02 tester for color fastness and light stability evaluation.
Air Cooled Light Fastness Tester Standards:
GB/T 8427, GB/T 8430, GB/T 14576, GB/T 15102, GB/T 15104, ISO 105-B02/B04, ISO 105-B10(requires irradiance meter 300-400nm), ISO 12040, ISO 11341, ATCC TM 16, AATCC 169, ASTM D5010, ASTM D4303,1, JIS L0843, M&S C09/C09A
Water cooled light fastness tester is widely use in solar radiation simulation light color fastness, light perspiration color fastness and light aging test of rubber, plastic, coating, petrochemical industry, car, textile industrial products and other materials. It through simulate illumination, exposed to the rain, mist spraying, alternating light and dark, temperature and humidity climates conditions to test material products’ color fading, aging, transmittance, peeling, hardness, softness and other performance changes.
This Xenon Weatherometer complies with ASTM G155 Testing Machine requirements for xenon weathering.
Xenon Weatherometer Standards:
Overview of UV Accelerated Weathering Tester
Weather Tester Standards:
ASTM G154, ASTMD 4329, ASTM D4587, ISO 16474-3, ISO 4892-1, ISO 11507, ISO4892-3, ISO11997-2, GB/T16422.3, GB/T16585, GB/T 14522-93
Key Comparison: Xenon Arc Tester vs. UV Weathering Tester
|
Item
|
Xeno
n Arc Tester
|
UV Weath
eri
ng Tester
|
|
|
Air‑Cooled
Xenon Arc Tes
ter
|
Water‑Cooled
Xenon Arc Tester
|
||
|
Light Source
|
2500W xenon long arc
|
6000W water‑cooled xenon arc
|
UVA‑340 fluorescent UV
|
|
Chamber Temperature
|
25
-
5
5
℃
,
Resolution:
0
.1
℃
|
20
-
93
℃
±
1
℃
Resolu
tion:
0.1
℃
|
RT+
1
0℃~70℃
|
| BPT |
40 - 80 ℃ , A ccuracy : ± 2 ℃ ; Resolution: 0.1 ℃ |
35
-12
5
℃
±1
℃
|
30~80℃
|
|
Chamber Humidity
|
Light Cycle : 10 - 70 % RH ; Resolution: 0.1 RH % Dar k Cycle : 30 - 95 % RH ; Resolution: 0.1 RH % |
Light Cycle: 10-70%RH ; Resolution :0.1% Dark Cycle: 3 0-95%RH ; Resolution :0.1% |
≥75%RH
|
|
Irradiance Control Range
|
0.80 - 1.20 W/m ² @420nm (Option: 340nm , 420nm , 300-400nm or 28 0-800nm ) A ccuracy : ± 0.02W/m ² @420nm |
0.6 ~ 3.08W/m²@420nm (Option: 340nm , 420nm , 300-400nm and 300-800nm) Accuracy: ± 0.02W/m²@420nm. |
0.
3
~1.3W/m2
,
(
Set irradiance on th
e touch screen
)
|
|
Spectrum
|
Full solar spectrum simulation (multi-band optional: 340nm, 420nm, 300–800nm) |
Highly accurate full solar spectrum simulation
|
UV spectrum only (accelerated aging)
|
|
Cooling
|
Air cooled
|
Water cooled
|
Air-cooled
|
|
Energy Efficiency
|
Low power
|
~3× air‑cooled
|
Moderate
|
|
Control
|
Closed‑loop irradiance
|
Higher power consumption for high-intensity testing
|
Moderate energy consumption
|
|
Spr
ay
Function
|
Yes
|
Yes
|
Yes
|
|
Body Material
|
Stainless steel
|
Stainless steel
|
SUS304# stainless steel (1.2mm)
|
Typical Applications
Conclusion
FAQ
Materials requiring full sunlight simulation: high‑grade coatings, automotive parts, textiles, outdoor plastics, architectural materials.
Per uno screening rapido del degrado indotto dai raggi UV, test di routine economici e materiali in cui è rilevante solo il danno da raggi UV.
I test UV spesso forniscono risultati visibili in pochi giorni; i test allo xeno possono richiedere settimane, a seconda del materiale e dello standard.
Lampade allo xeno raffreddate ad aria: fino a oltre 1000 ore; lampade allo xeno raffreddate ad acqua: lampade ad arco ad alta potenza e lunga durata; lampade UV: oltre 1600 ore.
SÌ.
Quale manutenzione è necessaria?
Ispezione periodica dei filtri/lampade, pulizia dell'impianto idrico (deionizzato e a spruzzo), calibrazione dell'irradianza e della temperatura e manutenzione del sistema di drenaggio.